DM100 User’s Manual Getting Started
2-8
2.7 Advanced Features
Measuring the Latency of Equipment
The DM100 can be used to measure the latency of other audio
equipment. At user-defined intervals, the DM100 generates a
short burst of white noise, then measures the time taken for the
signal to return.
Set the DM100 to measure latency by following these steps:
1. Go to the Gen Control / Output Mode setup menu, and
select Latency Test (see Section 4.2).
2. Connect the Digital Out from the DM100 to the
equipment to be measured.
3. Connect the output of the equipment to be measured to
the DM100’s Digital Input.
Different clock sources used for the generator can be set in the
I/O Control / Gen Clock Source setup menu (see Section 4.5).
The time interval between the noise bursts can be set in the
Gen Control / Noise Burst Rate setup menu (see Section 4.2).
To read the measured latency, press [Status] and navigate using
the [Down] button until Generator ENTER to View is
displayed. Press [Enter] to display the measured latency in either
AES samples or milliseconds (see Section 3.2.5).
Measuring Audio/Video Frequency Drift
The DM100 determine if an AES digital audio or reference
signal is frequency-locked to a video black burst reference
signal. The unit locks to the video reference, then measures the
frequency difference between the Digital In input and the video
reference.